• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFESCAN EUROPE GMBH OT ULTRA 2 METER; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE GMBH OT ULTRA 2 METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Model Number 021-098
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Fatigue (1849); Diaphoresis (2452); Increased Appetite (4570)
Date of Event 10/05/2020
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2020, the patient contacted lifescan (lfs) us, alleging that their onetouch ultra 2 meter was reading inaccurately high.This complaint was classified based on the customer care agent (cca) documentation.The patient reported that the alleged inaccuracy issue began at 01:00 on (b)(6) 2020.The patient claimed obtaining results of "556 mg/dl" (01:00, (b)(6) 2020), "181 m/dl" (02:04, (b)(6) 2020), "261 mg/dl" (08:43, 05 october 2020) and "204 mg/dl" (13:14, (b)(6) 2020) on the subject device.The patient manages their diabetes with oral medication ("glucozan, 5 mg", and "jardiance, 10 mg").The patient reported that she took "jardiance, 25 mg and, glucozan, 5 mg after each test taken" and that she stopped eating in response to the alleged elevated readings obtained on the subject device.The patient reported that at 02:00 on (b)(6) 2020, she became "weak", started to "sweat" and that she was "starving".The patient advised that she went to the hospital at 13:15 on (b)(6) 2020, where her blood glucose was reportedly tested at "55, 56 and 58 mg/dl" on the hospital device.The patient advised the cca that she was subsequently treated with orange juice and peanut butter crackers until her blood glucose reached "at least 90 mg/dl" during troubleshooting, the cca established that the unit of measure was set correctly on the subject device at the time of testing.The cca noted that the patient did not have control solution to perform a control solution test.Replacement products were sent to the patient.This complaint is being reported because the patient reportedly developed signs/symptoms suggestive of a serious injury adverse event and required treatment for an acute low blood glucose excursion from a health care professional after administering medication based on the alleged inaccurately high blood glucose readings obtained on the subject device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT ULTRA 2 METER
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug, 6300
SZ  6300
MDR Report Key10780008
Report Number2939301-2020-02749
Device Sequence Number11044385
Product Code NBW
UDI-Device Identifier00353885000277
UDI-Public00353885000277
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number021-098
Device Catalogue Number021-098
Device Lot Number4613180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/06/2020
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
-
-