Model Number N/A
Device Problems Failure to Zero (1683); Failure to Calibrate (2440)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2020
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted when additional information is provided.(b)(6).
 
Event Description
It was reported that the cardiosave intra aortic balloon pump (iabp) unit would not zero.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Event Description
It was reported that the cardiosave intra aortic balloon pump (iabp) unit would not zero.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, e1(event site email), e2, e3, g4, g7, h2, h6, h10, h11.Corrected fields: d5, h6(device codes).A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.The fse performed all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that the cardiosave intra aortic balloon pump (iabp) unit would not zero.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key10780261
MDR Text Key219372132
Report Number2249723-2020-01814
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/12/2020
Initial Date FDA Received11/03/2020
Supplement Dates Manufacturer Received11/11/2020
01/08/2021
Supplement Dates FDA Received12/05/2020
02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-