SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH CONNECT; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 66801791
Medical Device Problem Code Failure to Charge (1085)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/05/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the renasys touch device does not charge the battery when placed on the patient and switched on.It charges only after being switched off for a long time.Therapy continued with a back-up devices with no delay.No patient harm reported.
 
Additional Manufacturer Narrative
H10.The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH CONNECT
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10781547
Report Number8043484-2020-03596
Device Sequence Number11686023
Product Code OMP
UDI-Device Identifier05000223490782
UDI-Public05000223490782
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181822
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66801791
Device Catalogue Number66801791
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer01/11/2021
Initial Report FDA Received Date11/03/2020
Supplement Report FDA Received Date01/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-