• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN 6 DEG VAL COL 6MM OFF S3-8; PROSTH, KNEE, PATELLO/FEM, SEMI-CONS, UNCEM, POROUS, COATED, POL/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN 6 DEG VAL COL 6MM OFF S3-8; PROSTH, KNEE, PATELLO/FEM, SEMI-CONS, UNCEM, POROUS, COATED, POL/METAL/POLY Back to Search Results
Model Number 71434407
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/07/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a revision surgery the valgus collet 6mm was seized and would not rotate on the reamer shaft.The procedure was finished using a smith and nephew backup device.There was a delay reported of less than or equal to 30 minutes.No patient harm was reported.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device revealed nicks and deep scuffs on device.The device shows signs of extensive use.A functional evaluation of the returned device confirmed the stated failure mode.The rotating function of the device seized, rendering it inoperable.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN 6 DEG VAL COL 6MM OFF S3-8
Common Device Name
PROSTH, KNEE, PATELLO/FEM, SEMI-CONS, UNCEM, POROUS, COATED, POL/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10782147
Report Number1020279-2020-06103
Device Sequence Number19776440
Product Code MBH
UDI-Device Identifier03596010549150
UDI-Public03596010549150
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71434407
Device Catalogue Number71434407
Initial Date Received by Manufacturer 10/07/2020
Supplement Date Received by Manufacturer05/22/2021
Initial Report FDA Received Date11/03/2020
Supplement Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-