| Model Number |
368056 |
| Medical Device Problem Codes |
Short Fill (1575); Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported with the use of the bd vacutainer® lh 68 i.U.Plus blood collection tubes experienced under-fill or low draw of a tube with blood.The following information was provided by the initial reporter: the customer stated "the tube is under-filling.".
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d.10.Device available for eval?: yes d.10.Returned to manufacturer on: 10/23/2020 h.6.Investigation: bd received 10 samples for investigation.The samples were evaluated by draw water testing and the indicated failure mode for underfill with the incident lot was not observed.Additionally, 10 retention samples from bd inventory were evaluated by draw water testing and no issues were observed relating to underfill as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.Bd was not able to identifiy a root cause for the indicated failure mode.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported with the use of the bd vacutainer® lh 68 i.U.Plus blood collection tubes experienced under-fill or low draw of a tube with blood.The following information was provided by the initial reporter: the customer stated "the tube is under-filling.".
|
| |
|