| Model Number |
DL-330-10-K |
| Medical Device Problem Codes |
High impedance (1291); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Review of the lead impedance and ecog data was performed and the data is consistent with a lead break.Neuropace received only the proximal end of the explanted lead for investigation.The portion of the lead received was received damaged and the coil was stretched.No wire breaks were found on the portion of the lead that was received, all 4 wires passed continuity test.Unable to determine root cause of the lead break.
|
| |
|
Event or Problem Description
|
|
A review of the recorded ecog data and impedance measurements on pdms confirmed high impedances and signal artifact on the left depth lead, suggestive of a lead break.It was unclear exactly when the changes occurred between february and june as the patient was not regularly using his remote monitor.The patient did not recall falling in between the time that the signal changes were noted on the depth lead.A lead replacement was performed and a portion of the explanted lead was discarded by the hospital.A new lead was implanted and connected to the rns neurostimulator without complications and is functioning as intended.
|
| |
|