• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE, 20F; FEEDING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE, 20F; FEEDING DEVICE Back to Search Results
Model Number 000720
Medical Device Problem Codes Material Rupture (1546); Device Dislodged or Dislocated (2923); Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 08/2022).
 
Event or Problem Description
It was reported that approximately 15 days post feeding tube placement, the catheter allegedly dislodged.It was further reported that the catheter fell out of the patient.There was no reported patient injury.
 
Event or Problem Description
It was reported that approximately fifteen days post feeding tube placement, the catheter allegedly was dislodged.It was further reported that the catheter fell out of the patient.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one tri-funnel replacement gastrostomy tube 20f was returned for evaluation.Gross, microscopic evaluation and functional testing were performed.The investigation is confirmed for tube dislodged, migration and balloon rupture issue as a rupture to the balloon was noted from the distal end of the device.The balloon was inflated with approximately 10ml of water and immediately deflated.A leak was observed from the rupture on the balloon.A definitive root cause could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2022), g3.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE, 20F
Common Device Name
FEEDING DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key10784384
Report Number3006260740-2020-20363
Device Sequence Number11712078
Product Code KNT
UDI-Device Identifier00801741037153
UDI-Public(01)00801741037153
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000720
Device Catalogue Number000720
Device Lot NumberNGDW3857
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Initial Date Received by Manufacturer 10/08/2020
Supplement Date Received by Manufacturer04/01/2021
Initial Report FDA Received Date11/04/2020
Supplement Report FDA Received Date04/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-