Product complaint # (b)(4).Date sent to the fda: 12/09/2020 additional information: d9 a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Investigation summary = > it was reported that the problem of pulling off suture needle had happened in surgical treatment of congenital heart disease in children surgery.An opened box with unopened samples of product coded8413, lot # kcm981 were received for evaluation.During the visual inspection of unopened samples, no defects were found on the packages.The samples were opened, and swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|