| Model Number |
N/A |
| Medical Device Problem Codes |
Unstable (1667); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
09/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 00771100700-femoral stem 12/14 neck taper plasma sprayed press-fit cementless size 7.5 standard offset-64496778.00885100832-neutral liner 32 mm-63468929.00625006530-bone scr 6.5x30 self-tap-j6779178.00877503204- biolox⮠delta, ceramic femoral head, xl, 㸠32/+7, taper 12/14- 2993195.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2020 - 03728.0001822565 - 2020 - 03730.Customer has indicated that the product will not be returned to zimmer biomet for investigation, product location unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported patient underwent a left hip revision approximately 1-month post implantation due to subluxation, malposition, and instability.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: g4; h2; h3; h6.Reported event was confirmed via operative notes and x-rays reviewed by a healthcare professional.Review of the device history record for the shell identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|