| Model Number |
4FC12 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the 4fc12 sheath with lot 0010066558 was returned and analyzed.Visual inspection of the sheath showed the shaft was intact with no apparent issues.Performance testing with the leak tester revealed the pressure decay was in an acceptable range.The shaft and valve were leak tight with no apparent issues.Several flushes and aspirations were performed without issues while the balloon catheter was inserted.There was no blockage along the shaft and the shaft and side port were leak tight.In conclusion, the reported air ingress issue was not confirmed through testing.The sheath passed the returned product inspection as per specification.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cryo ablation procedure, air bubbles were noted after the balloon catheter was inserted in the sheath.The balloon catheter was re-inserted into the sheath without resolving the issue.The sheath was then replaced which resolved the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
| |
|