| Model Number |
6L45-21 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely elevated direct bilirubin results when compare to total bilirubin on architect c16000 processing module for one patient.The results provided were: total bilirubin on architect=1.7 mg/dl/ repeated on same analyzer=1.6 mg/dl/ repeated on alinity=2.07 mg/dl/ repeated on siemens exl=2.6 mg/dl laboratory reference range for architect total bilirubin=0.2 mg/dl to 1.2 mg/dl laboratory reference range for exl total bilirubin=0.2 mg/dl to 1.0 mg/dl there was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided: describe event or problem b5: event date and sid provided.The customer observed falsely elevated direct bilirubin results when compare to total bilirubin on architect c16000 processing module for one patient.The results provided were: on (b)(6) 2020 (b)(6) total bilirubin on architect=1.7 mg/dl/ repeated on same analyzer=1.6 mg/dl/ repeated on alinity=2.07 mg/dl/ repeated on siemens exl=2.6 mg/dl laboratory reference range for architect total bilirubin=0.2 mg/dl to 1.2 mg/dl laboratory reference range for exl total bilirubin=0.2 mg/dl to 1.0 mg/dl there was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of complaints for lot 56326uq11 identified only one other complaint for total bilirubin results.The trending report review determined that there are no trends for the product for the complaint issue.Return testing was not completed as returns were not available.Per product labeling, the presence of fibrin, bubbles or particulate matter in the sample may cause an erroneous result.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of labeling concluded that the issue is sufficiently addressed in the labeling.Based on the investigation no product deficiency was identified for the architect total bilirubin reagent, lot 56326uq11.
|
| |
|