| Model Number |
662878 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported while testing for sars cov-2 a false positive result was obtained.The patient was retested on the veritor and the result was negative.The customer stated they are testing asymptomatic patients.This test is not intended for use on asymptomatic patients, and was therefore used off label.There was no indication that results were reported out, and there was no report of patient impact.Eua#: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported while testing for sars cov-2 a false positive result was obtained.The patient was retested on the veritor and the result was negative.The customer stated they are testing asymptomatic patients.This test is not intended for use on asymptomatic patients and was therefore used off label.There was no indication that results were reported out and there was no report of patient impact.Eua#: eua201889.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: bd has received several customer complaints for false positive results, when using bd sars-cov-2 reagents for bd veritor¿ system.The current investigation concerns multiple lots of bd sars-cov-2 reagents for bd veritor¿ system.Bd takes a systematic approach to investigating false positive complaints that are received.This approach involves review of manufacturing batch history records for batch number 0246062 testing of retention samples of batch number 0246062, and testing of customer returned samples if applicable.The investigation did not find a root cause for the false positive results that were observed.The root cause is under investigation and will be documented in our quality system.Bd cannot confirm the complaint based on the investigation that was performed.A corrective and preventive action (capa#1878253) is already initiated to investigate the root cause and some mitigation actions are already being addressed.H3 other text : see h.10.
|
| |
|