MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM COLLAR LAT. TI/HA 4; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENT, METAL/POLY, NON-POROUS, CA-PHOSPH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM COLLAR LAT. TI/HA 4; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENT, METAL/POLY, NON-POROUS, CA-PHOSPH Back to Search Results
Model Number 75018415
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 10/13/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent revision surgery due to constant dislocations.The surgeon attempted to replace the liner and add 3 screws and a r3 liner to obtain stability but was not achieved.After this the cup was removed and also the stem.The final result was successful and the wound was closed.
 
Additional Manufacturer Narrative
Additional information provided in: d10 (additional concomitant device was added) and ( health effect - impact code and component code) results of investigation: it was reported that the patient underwent revision surgery due to constant dislocations.A polarstem collar lat.Ti/ha 4 became loose during the case and was revised.The device, which intent use is in treatment, was not received for investigation.Furthermore no clinical documenation got available.A design history review was performed.No deviations could be found in the corresponding batch record review.For the specific batch number no other complaint was found.There is no indication that the device failed in any way.The root cause for this failure remains undetermined.A review of the risk management documentation verifies the failure mode and severity of the reported issue.The current instruction for use for hip implants 12.23 ed.05/16 lists dislocation as a known possible side effect resulting from a hip arthroplasty.Internal complaint reference case-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSTEM COLLAR LAT. TI/HA 4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENT, METAL/POLY, NON-POROUS, CA-PHOSPH
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
MDR Report Key10788682
Report Number9613369-2020-00227
Device Sequence Number19560282
Product Code MEH
UDI-Device Identifier07611996103526
UDI-Public07611996103526
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K143739
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number75018415
Device Catalogue Number75018415
Device Lot NumberB1925014
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/13/2020
Supplement Date Received by Manufacturer12/18/2020
Initial Report FDA Received Date11/04/2020
Supplement Report FDA Received Date12/23/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEAD 71343608, LOT 19MM12098; REF INTERFIT THRD HOLE COVER (71330001)
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-