• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided; best estimate is between (b)(6).(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a tecnis symfony intraocular lens (iol) was explanted from the patient's right (od) eye because she was not happy with far vision.The replacement lens is a non-johnson and johnson lens.Reportedly, patient is doing good post surgery.No additional information provided.
 
Additional Manufacturer Narrative
Section d10: device available for evaluation: yes.Section d10: returned to manufacturer on: 10/21/2020 , section h3: device returned to manufacturer: yes.Device evaluation: the complaint lens was received stuck to a completed jjsv shipping guide.Additional documentation was received as well.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptic, and that the lens was received cut in half, which is consistent with a lens handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10788761
Report Number2648035-2020-00816
Device Sequence Number2904208
Product Code POE
UDI-Device Identifier05050474579101
UDI-Public(01)05050474579101(17)230807
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/07/2020
1 Device was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/07/2023
Device Model NumberZXR00
Device Catalogue NumberZXR00U0205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Initial Date Received by Manufacturer 10/08/2020
Supplement Date Received by Manufacturer11/09/2020
Initial Report FDA Received Date11/04/2020
Supplement Report FDA Received Date12/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
-
-