Catalog Number 0117060
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As reported, a foreign material was found on the perfix light plug.The material was removed from the device and the device was implanted into the patient.The material removed was returned for evaluation.The evaluation confirmed the material to be a piece of the monofilament used when stitching the petals of the perfix light plug together.There was no alleged detachment of the plug during use, indicating that the material found is residual from the manufacturing process likely from the when the thread is cut.Based on the investigation and returned sample evaluation, the root cause is determined to be manufacturing related.A review of the manufacturing records was performed and found that the lot was manufactured to specification.To date, this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in february, 2020.
 
Event or Problem Description
As reported, it was identified that while opening the package on (b)(6) 2020, a foreign material like frayed thread was found on the perfix light plug.The device was used after removing the foreign material.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFIX PLUG LIGHT
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
MDR Report Key10790538
Report Number1213643-2020-20033
Device Sequence Number11034875
Product Code FTL
UDI-Device Identifier00801741030956
UDI-Public(01)00801741030956
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K092032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/28/2024
Device Catalogue Number0117060
Device Lot NumberHUEN0049
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/19/2020
Initial Report FDA Received Date11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-