| Catalog Number |
0117060 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As reported, a foreign material was found on the perfix light plug.The material was removed from the device and the device was implanted into the patient.The material removed was returned for evaluation.The evaluation confirmed the material to be a piece of the monofilament used when stitching the petals of the perfix light plug together.There was no alleged detachment of the plug during use, indicating that the material found is residual from the manufacturing process likely from the when the thread is cut.Based on the investigation and returned sample evaluation, the root cause is determined to be manufacturing related.A review of the manufacturing records was performed and found that the lot was manufactured to specification.To date, this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in february, 2020.
|
| |
|
Event or Problem Description
|
|
As reported, it was identified that while opening the package on (b)(6) 2020, a foreign material like frayed thread was found on the perfix light plug.The device was used after removing the foreign material.There was no reported patient injury.
|
| |
|