| Model Number |
MC9079 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample is available for evaluation.It is yet to be received.
|
| |
|
Event or Problem Description
|
|
The customer reported a 17" ext set w/0.2 micron filter, microclave® clear, clamp, rotating luer that had air in the iv tubing when changing out the lines.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Three (3) new list# mc9079, extension sets (lot# 4915987) were received on november 9, 2020 and visually inspected.As received, no damage or anomalies were seen on any of the returned sets.No used product or mating devices were returned.Each extension set was primed with water at gravity pressure and then leak tested according to product performance specifications.No leaks or air bubbles were seen.The reported complaint could not be replicated or confirmed.A device history review for lot# 4915987 and relevant commodities were reviewed, and no non-conformances were found that would have contributed to the reported complaint.
|
| |
|