BIOMET MICROFIXATION MODIFIED BECK ELEVATOR; ELEVATOR, SURGICAL, DENTAL Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported the instrument broke at the tip during an elevation.No adverse events have been reported as a result of the malfunction.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, g3, g6, h2, h3, h6, h10 visual examination of the returned product confirmed the identity of the product and also confirmed the reported fracture at the tip of the instrument.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The device was returned to the supplier for evaluation.The supplier tested the hardness and determined the value was within the specifications.The supplier also stated the fracture point shows that the instrument was not used according to the purpose and the instrument must not be used as a level.Due to the special shape, an over-strengthening of the tip can cause breakage.No medical records were supplied with the complaint, therefore no comments can be made regarding the surgical technique.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODIFIED BECK ELEVATOR
Common Device Name
ELEVATOR, SURGICAL, DENTAL
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key10791401
Report Number0001032347-2020-00560
Device Sequence Number1088248
Product Code EMJ
UDI-Device Identifier00841036285302
UDI-Public00841036285302
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberSP-2359
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/26/2020
Supplement Date Received by Manufacturer06/16/2021
Initial Report FDA Received Date11/05/2020
Supplement Report FDA Received Date06/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-