| Model Number |
J488G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.- what tissue was being approximated or sutured when it broke? a manufacturing record evaluation was performed for the finished device lot pe6800, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that an animal underwent an eye procedure on (b)(6) 2020; and a suture was used.During the procedure, the suture broke apart.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent to the fda: 12/15/2020.Additional h6 component code: g07002 - conforming device.Additional h3 investigation summary: it was reported by the distributor that during a vet eye procedure that the suture broke apart.The tensile strength seemed brittle.An opened box with nine unopened samples of product code j488g, lot # pe6800 were received for evaluation.During the visual inspection of nine unopened samples, no defects were found on the package.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The device performed without any defect noted and the actual complaint sample was not returned for analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Date sent to the fda: 12/01/2020.Additional information was requested, and the following was obtained: what tissue was being approximated or sutured when it broke? we were suturing gingiva when it broke.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|