| Model Number |
32899 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for analysis.Visual inspection was performed and the returned device was found with the shaft middle section detached/separated.The flexible cannula distal section was returned loaded in the catheter distal section.The catheter and flexible cannula show evidence of cut at section detached.The flexible cannula was unloaded form the catheter without issues.Functional inspection was performed and a 0.038" mandrel was inserted through the catheter and no resistance was felt.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 27oct2020.It was reported that the flexible cannula became stuck and broke inside the catheter.An 8.3/26 expel nephroureteral stent system with twist-loc hub was selected for use.During the procedure, it was noted that the plastic stylet would get stuck in the drainage catheter with the wire in place.Then, it was noted that the soft plastic stiffener broke off in the drainage catheter.The physician had to remove everything from the patient and start over.The procedure was completed with a non-boston scientific product.No patient complications were reported.However, device analysis revealed that the shaft middle section was detached/separated.
|
| |
|