MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIG RP AOX CRV INS SZ4 10; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIG RP AOX CRV INS SZ4 10; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT Back to Search Results
Model Number 1961-92-041
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical notification received for revision of right total knee to address instability date of implantation: (b)(6) 2020.Date of revision: (b)(6) 2020 (left knee).Treatment: revision of femur and tibial insert components; tibial tray and patella were retained.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).No device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG RP AOX CRV INS SZ4 10
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10791946
Report Number1818910-2020-23993
Device Sequence Number2875440
Product Code NJL
UDI-Device Identifier10603295076209
UDI-Public10603295076209
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1961-92-041
Device Catalogue Number196192041
Device Lot Number9455176
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/22/2020
Supplement Date Received by Manufacturer12/15/2020
Initial Report FDA Received Date11/05/2020
Supplement Report FDA Received Date12/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MBT CEM KEEL TIB TRAY SZ4; PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG C/R POROCOAT FEM LT SZ 4; SIG RP AOX CRV INS SZ4 10; UNKNOWN BONE CEMENT; UNKNOWN BONE CEMENT; MBT CEM KEEL TIB TRAY SZ4; PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG C/R POROCOAT FEM LT SZ 4; SIG RP AOX CRV INS SZ4 10; UNKNOWN BONE CEMENT; UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient Weight91
-
-