BARD PERIPHERAL VASCULAR, INC. DENALI JUGULAR SYSTEM; VENA CAVA FILTER Back to Search Results
Model Number DL900J
Device Problems Premature Activation (1484); Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2020
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo is provided for a review.The investigation of the reported event is currently underway.(expiry date: 11/2022).
 
Event Description
It was reported that prior to a filter placement procedure, the filter was allegedly found to be partially deployed.The procedure was completed using another device.There was no patient contact.
 
Manufacturer Narrative
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported for this product/lot number combination.Therefore, a device history record review is not required.Investigation summary: the device was not returned for evaluation.Two photos were provided, however no evidence of partial deployment could be confirmed by the provided photos.Therefore, the investigation will remain inconclusive for the reported deployment issue.The definitive root cause could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 11/2022).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that prior to a filter placement procedure, the filter was allegedly found to be partially deployed.The procedure was completed using another device.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI JUGULAR SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key10792252
MDR Text Key214785748
Report Number2020394-2020-20345
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00801741040795
UDI-Public(01)00801741040795
Combination Product (y/n)N
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDL900J
Device Catalogue NumberDL900J
Device Lot NumberGFDX1885
Was Device Available for Evaluation? No
Date Manufacturer Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-