| Model Number |
RBY4C1650-A |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2020-01891.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the renal artery using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, the physician placed a ruby coil into the aneurysm; however, the physician was not satisfied with the placement of the ruby coil.While retracting the ruby coil to reposition it, the physician experienced resistance and subsequently, the ruby coil became stuck within the lantern.Therefore, the physician removed all devices from the patient and then removed the ruby coil from the lantern.The physician then placed two ruby coils into the aneurysm using the lantern.While implanting the last ruby coil (f93620) inside the aneurysm, the ruby coil unintentionally detached and subsequently, part of the ruby coil was in the renal artery.Therefore, the physician placed a stent to hold part of the ruby coil within the vessel wall.The procedure ended at this point.There was no report of an adverse effect to the patient.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the renal artery using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, while advancing a ruby coil through the lantern, the operator kinked the ruby coil.Subsequently, the lantern and ruby coil were removed from the patient.The ruby coil was removed from the lantern and the microcatheter was re-introduced into the patient.Then, the physician placed a ruby coil into the aneurysm; however, the physician was not satisfied with the placement of the ruby coil.While retracting the ruby coil to reposition it, the physician experienced resistance and the ruby coil became stuck within the lantern.Therefore, the physician removed all devices from the patient and then removed the ruby coil from the lantern.The physician then placed two ruby coils into the aneurysm using the lantern.While implanting the last ruby coil (f93620) inside the aneurysm, the ruby coil unintentionally detached and subsequently, part of the ruby coil was in the renal artery.Therefore, the physician placed a stent to hold part of the ruby coil within the vessel wall.The procedure ended at this point.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note that the following section is being updated based on additional information provided by a penumbra sales representative on 11/25/2020: section b.Box 5.Describe event or problem.Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked in multiple locations throughout its length.The pusher assembly was fractured approximately 129.0 cm from the proximal end.The embolization coil was intact with its pusher assembly and had offset coil winds.Conclusions: evaluation of the first returned ruby coil revealed a kink and fractured pusher assembly.If the ruby coil is forcefully advanced against resistance, damage such as a kink and subsequently fracture may occur.Further evaluation revealed additional kinks throughout the length of the pusher assembly.This damage may have been incidental to the reported complaint, and likely occurred during packaging of the device for return.Evaluation of the second returned ruby coil confirmed that the pusher assembly was kinked.If the ruby advanced against resistance or otherwise mishandled, the pusher assembly can be kinked.Further evaluation revealed the embolization coil was detached.This damage likely occurred during removal from the microcatheter on the back table and was incidental to the reported complaint.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2020-01891; 3005168196-2020-02353.
|
| |
|