| Model Number |
031328-10 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product has not yet been returned for investigation.
|
| |
|
Event or Problem Description
|
|
As per a manufacturer incident report we received from the factory in (b)(4): after mounting the tubing on the arthropump, leakage of physiological serum observed.Consequent leakage after insertion of the arthroscope into the joint concerned (pump blockage); stopped the operation and used a different tubing.
|
| |
|
Additional Manufacturer Narrative
|
|
The tubing set was returned for investigation on november 09, 2020.Upon evaluation, it could be confirmed that there is leak at the connection area.The concrete root cause could not be identified, however the leakage at the connection area is potentially due to gluing or mechanical damage such as bend this is the first case in total (b)(4) product sold worldwide.
|
| |
|