| Model Number |
1221-32-154 |
| Medical Device Problem Codes |
Fracture (1260); Device Dislodged or Dislocated (2923); Migration (4003)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Inflammation (1932); Joint Dislocation (2374); No Code Available (3191)
|
| Date of Event |
10/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Hip revision surgery found a finding in the acetabulum; patient who had had one of our pinnacle + corail prostheses placed two years ago; the insert is broken and the femoral head of the prosthesis is out of the socket causing metallosis of the entire area.Product: 122132154, lot: not reported.Product: 136522000, lot: not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Review of the attached image could not confirm a dislocation event but could confirm the liner has fractured.A root cause could not be determined.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|