Model Number 62-3411L
Medical Device Problem Codes Loss of or Failure to Bond (1068); Fracture (1260); Loss of Osseointegration (2408)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); No Code Available (3191); Insufficient Information (4580)
Date of Event 10/21/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient was revised due to loosening of the srom femur at the bone to implant interface.Post of srom hinge femoral component broke in-situ and part of post was stuck in universal femoral sleeve.Surgeon believes it broke under pressure.Surgeon had to remove femoral sleeve and press fit stem.Femur was revised.Doi: (b)(6) 2017.Dor: (b)(6) 2020; left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : dhr reviewed no deviations or nonconformance's were noted.H6 clinical code: removed the code for inadequate osteointegration and replaced with insufficient information (e2401).H6 medical device problem code: removed code for osseointegration and replaced with loss of or failure to bond (a040102).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SROM NRHFEM W/PIN SM LFT 66X62
Common Device Name
LPS AND S-ROM : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10792900
Report Number1818910-2020-24032
Device Sequence Number2988431
Product Code KRO
UDI-Device Identifier10603295179481
UDI-Public10603295179481
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K896048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number62-3411L
Device Catalogue Number623411L
Device Lot NumberC88168
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/21/2020
Supplement Date Received by Manufacturer12/14/2020
Initial Report FDA Received Date11/05/2020
Supplement Report FDA Received Date12/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LPS UNIV TIB HIN INS SM 12MM; UNIVERSAL FEM SLV FUL POR 46MM; UNIVERSAL STEM 150X20MM FLUTED; LPS UNIV TIB HIN INS SM 12MM; UNIVERSAL FEM SLV FUL POR 46MM; UNIVERSAL STEM 150X20MM FLUTED
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-