| Model Number |
24672 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.The 50-60% stenosed target lesion was located in the non-tortuous and moderately calcified superficial femoral artery.A 6.0 x 150, 135cm mustang balloon catheter was advanced for dilation.However, after deflation, the balloon refused to exit and became stuck over the wire.The device was removed with the whole system and completed the procedure with another of same device.There were no patient complications reported and the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: mustang device 6x15x135cm was received for analysis.As per the mustang direction for use the catheter is compatible with a 0.035in (0.89 mm) guidewire.The investigator successfully loaded a boston scientific 0.035in guidewire through the mustang device and removed it with no resistance encountered.No issues were identified with the device that could potentially have contributed to the complaint incident.A visual and tactile examination identified that the balloon had been subject to positive pressure.No damage or issues were noted which may have potentially contributed to the complaint incident.A visual and tactile examination identified no issues with the shaft of the device.No issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.The 50-60% stenosed target lesion was located in the non-tortuous and moderately calcified superficial femoral artery.A 6.0 x 150, 135cm mustang balloon catheter was advanced for dilation.However, after deflation, the balloon refused to exit and became stuck over the wire.The device was removed with the whole system and completed the procedure with another of same device.There were no patient complications reported and the patient was stable.
|
| |
|
Search Alerts/Recalls
|