• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM LCP RECONSTRUCTION PLATE 12 HOLES/168MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM LCP RECONSTRUCTION PLATE 12 HOLES/168MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 245.121
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.(b)(4).Without a lot number the device history records review could not be completed.Product was not returned.The instrument(s) was not returned and instead the investigation will be done based on the supplied image(s).The image(s) was reviewed and the complaint condition could not be confirmed.There was no allegation against the device, no malfunction, damage, or defects on the x-rays.As the instrument(s) was not returned an as received condition, dimensional inspection, material or drawing reviews are not applicable.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent a revision procedure with anatomic reduction for the 3 cortex screw due to malalignment of distal ulna fracture.The proximal screws were retained.Distal screw removed as part of a distal fractur re-alignment.Once reduction was restored, new distal screws were implanted.The procedure was successfully completed with no surgical delay.The patient's status is unknown.This report is for 1 3.5mm lcp reconstruction plate 12 holes/168mm.This is report 4 of 4 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5MM LCP RECONSTRUCTION PLATE 12 HOLES/168MM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10793520
Report Number2939274-2020-04983
Device Sequence Number3334445
Product Code KTT
UDI-Device Identifier10886982169686
UDI-Public(01)10886982169686
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K000684
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number245.121
Device Catalogue Number245.121
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/09/2020
Initial Report FDA Received Date11/05/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-