|
The reported event was inconclusive, as no sample returned for evaluation.A potential root cause for this failure mode could be due to "bad fit with the shaft." the device used for the treatment, though it was unknown whether the device related to the reported event or met specifications.The device was not returned for evaluation.The lot number was unknown, therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.".
|