MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Electromagnetic Interference (1194); Overheating of Device (1437); Vibration (1674)
|
Patient Problems
Pain (1994); Therapeutic Effects, Unexpected (2099); Burning Sensation (2146); Therapeutic Response, Decreased (2271); Complaint, Ill-Defined (2331); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 10/21/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that caller reports patient had 4 different mri scans done on (b)(6) 2020.Two scans were on patient's shoulders, one scan on patient's cervical spine, and one scan on patient's lumbar.Caller reports each scans were 30 mins and 90 mins between each scan.Caller reports since the mri scan, patient feels vibration at the ins pocket site and warmth across left side of the spine.Caller reports patient had programmed his device into mri mode and full body was seen.Caller reports the vibration and warmth latest until the mri tech decreased the mri machine.Caller does not know what was decreased.Caller reports since patient had mri scans on (b)(6) 2020, reports his therapy is different, not the same relief as before the mri scans.Caller reports prior to mri scans, he was getting great pain relief and now the stimulator is not helping, no pain relief.Caller reports patient is getting shooting pain on both his legs since mri scans.Caller reports ins interrogated normal, normal impedance.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the manufacturer representative (rep) reporting that the cause of the change in therapy/stimulation after the mri scans was not determined.The battery was being replaced with a new battery and the battery related to this complaint was being sent back for analysis.It was unknown if the issue was resolved at the time of the report.It was indicated that the provided information had been confirmed with the physician.No further complications were reported/anticipated.
|
|
Manufacturer Narrative
|
Product analysis #260042775:analysis information -- 2021-05-21 14:20:19 cst pli# 10 product id# 97715 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection, output and telemetry testing, and functional testing.The implantable neurostimulator (ins) passed functional testing.H3: analysis of the 97715 ins (s/n: (b)(6)) showed that the ins was functionally okay with insignificant anomalies.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|