| Model Number |
0502904070 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was lack of image.
|
| |
|
Event or Problem Description
|
|
It was reported that there was lack of image.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: no image.Confirmed failure: outer tube damaged (bent, dented),loose prism,damaged distal cover glass.Probable root cause: incorrectly assembled optical train; damage to optical train; debris in optical train; end of life wear-out; shipping damage; use error.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|