• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS SM 2.5MM TAPERED HEX DR SHAFT-SHORT; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS SM 2.5MM TAPERED HEX DR SHAFT-SHORT; SCREWDRIVER Back to Search Results
Model Number 71175073
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/15/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the evos sm 2.5mm tapered hex driver shaft-short has become worn out and is having difficulty capturing the screws.There was no case involved.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The hex end of the device is dull and worn rendering it inoperable.The device was manufactured 2019 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SM 2.5MM TAPERED HEX DR SHAFT-SHORT
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10795506
Report Number1020279-2020-06233
Device Sequence Number1126194
Product Code HXX
UDI-Device Identifier00885556636053
UDI-Public00885556636053
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175073
Device Catalogue Number71175073
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/15/2020
Supplement Date Received by Manufacturer10/15/2020
Initial Report FDA Received Date11/05/2020
Supplement Report FDA Received Date04/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-