COVIDIEN LP LLC NORTH HAVEN SURGIPRO LL; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
|
Back to Search Results |
|
| Model Number |
VP-901-MX |
| Medical Device Problem Codes |
Use of Device Problem (1670); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a coronary bypass, when the central anastomosis was performed, the needle came off the thread when holding the needle.Also, it was difficult to slip the string downward than usual.After that, a product with the same lot number was opened and the operation was completed successfully, but it was required to check the needle and thread coating.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection found six sutures and five needles.Three of the sutures were partial.Two of the sutures remained wound inside the retainer and their needles were properly parked in the foam of the retainer.Two needles were attached to each suture, one needle at each end.The last suture had only one needle attached with the other end missing a needle.Microscopic inspection of the suture with the missing needle noted that the end was deformed and flattened.Normal crimp and swage marks were noted on the attached needles.A tactile and microscopic inspection of the wound sutures was performed with no abnormalities.Microscopic inspection of the properly parked needles noted no abnormalities.The sutures were inspected for end to end for any fraying or discoloration in the surface of the suture and noted no abnormalities.Functional testing noted that a needle attachment test was performed on the wound samples, and the results met the specifications for this product.It was reported that the needle unexpectedly separated from the thread.The reported issue was confirmed.The most likely cause could not be established from the information available.It was also reported that the suture is too difficult to slip.The reported issue could not be confirmed.The most likely cause could not be identified because no rel ated problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open abdominal surgery, when the central anastomosis was performed, the needle came off the thread when holding the needle.Also, it was difficult to slip the string downward than usual.After that, a product with the same lot number was opened and the operation was completed successfully, but it was required to check the needle and thread coating.There was no patient injury.
|
| |
|