DEPUY SYNTHES PRODUCTS LLC UNIVERSAL BATTERY CHARGER II; BATTERY, REPLACEMENT, RECHARGEABLE
|
Back to Search Results |
|
| Catalog Number |
05.001.204 |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The reporters phone number was not provided.Concomitant medical products and therapy dates: battery devices, (b)(6) 2020.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from (b)(6) that the universal battery charger device showed fumes and sparks while charging the battery devices.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device did not function and there was component damage.It was further determined that the device failed pretest for check software revision and power-on test.The assignable root cause was determined to be due to component failure from normal wear.
|
| |
|
Search Alerts/Recalls
|
|
|