| Model Number |
MMT-723RNAS |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump passed the displacement test.Unit received with motor error alarm during the basic occlusion test due to loose/flush drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test due to motor error alarm.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had motor error alarm.Customer reported that the drive support cap was correct.Customer reported that the insulin pump was no dropped or bumped.Customer was able to successfully clear the alarm.Customer was able to complete pump rewind.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|