HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSIST) BYPASS
|
Back to Search Results |
|
Device Problems
Increase in Suction (1604); Obstruction of Flow (2423)
|
Patient Problems
Thrombosis (2100); Thrombosis/Thrombus (4440)
|
Event Date 01/01/2012 |
Event Type
Injury
|
Manufacturer Narrative
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac trans plantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Medtronic was made aware of this event through a search of literature publications.It was not possible to ascertain specific device information from the literature publication or to match the event with previously reported events.This information is based entirely on journal literature.This event occurred outside the u.S.All information provided is included in this report.Patient information is limited due to confidentiality concerns.Multiple patients were noted in the article; however, a one to one correlation could not be made with unique device serial numbers.The baseline gender/age of the patients represented in the article is male/54 years old.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.Request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: discontinuation of heartware rvad support without device removal in chronic bivad patients.American society for artificial internal organs journal, january-february 2012; 58(1):15-8.Doi: 10.1097/mat.0b013e3182376b7b.Additional information has been requested regarding the cause of the event, but it was not available at the time of this report.If additional information is received, the event will be updated and a supplemental report will be sent.Investigation of this event is pending and a supplemental report will be sent upon its completion.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
A journal article was reviewed which contained information regarding ventricular assist devices (vads).The article discussed in situ deactivation of right vads (rvads) for myocardial recovery in three patients using biventricular vads (bivads) for support.Multiple patients were noted in the article; however, a one-to-one correlation could not be made with unique device serial numbers.Two patients on bivad support, who gained right ventricular recovery, started experiencing subsequent suction events, then rvad thromboses.The rvad pumps were stopped, then surgical excision and removal of the rvad drivelines was performed.The two patients remain on left vad (lvad) support with the rvad pumps deactivated in situ.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for investigation completion.Product event summary: pumps with unknown serial numbers were not returned for evaluation.Log file analysis could not be performed since log files covering the reported event date were not available for analysis.As a result, the reported suction events could not be confirmed.Review of the sterility certificates could not be conducted since the serial numbers are unknown.Based on the investigation conducted, there is no evidence to suggest that a device malfunction caused or contributed to the reported events.Based on the risk documentation, possible causes of the reported suction events may be attributed to multiple factors including but not limited to thrombus at the inflow cannula, poor vad filling, and/or inappropriate pump rotational speed.Possible clinical factors that may have contributed to these events include the patients¿ pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patients¿ complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to these events.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|
|
|