DEPUY ORTHOPAEDICS INC US PFC*SIGMA TC3 FEM LEFT SZ5; SIGMA REVISION IMPLANT : KNEE FEMORAL Back to Search Results
Model Number 96-0084
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/20/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Aseptic revision of left knee.Knee had been revised many years ago and the femoral component was unsupported by bone so in time the stress was all on the bolt and it failed.The femoral component was removed, the adaptor was removed from the well ingrown sleeve and an srom hinge femur was inserted into the femoral sleeve, a 12mm hinge insert was placed and the wound was closed.I was not provided with original date of surgery so i read the lot numbers as much as possible from the removed implants.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SIGMA REVISION IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10800853
Report Number1818910-2020-24067
Device Sequence Number1126254
Product Code JWH
UDI-Device Identifier10603295232513
UDI-Public10603295232513
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K952830
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number96-0084
Device Catalogue Number960084
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/20/2020
Supplement Date Received by Manufacturer12/07/2020
Initial Report FDA Received Date11/06/2020
Supplement Report FDA Received Date12/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SIG FEM ADAP +2/-2 OFFSET BOLT; SIGMA FEM ADAPTER 5 DEGREE; SIGMA TC3 GVF INS 5 10MM; SIG FEM ADAP +2/-2 OFFSET BOLT; SIGMA FEM ADAPTER 5 DEGREE; SIGMA TC3 GVF INS 5 10MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient Weight100
-
-