SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/10/2020
Event Type  malfunction  
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation. There was no relationship found between the returned device and the reported incident. A visual inspection was performed on the product and observed a broken lanyard. A functional evaluation revealed a handpiece stall error. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended. A complaint history review concluded this was a repeat issue. The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process. Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event. No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that during a procedure, the motor drive unit had a handpiece sensor fault. The procedure was completed with the same device and no significant delay or patient injuries were reported. Results of investigation have concluded that this unit displayed a hand piece stall error which makes it a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10801574
MDR Text Key215080898
Report Number1643264-2020-01715
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeCS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/06/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage

-
-