Information was received from the consumer via the manufacturer representative (rep) regarding a patient with an implantable neurostimulator (ins).It was reported that there were discharges.The device discharged six times a day.Factors that may have led or contributed to the issue was no event.Troubleshooting included checking impedances which were okay, changing programs, stopping the adaptive stimulation, and stopping stimulation where there were no more discharge.An appointment was scheduled with the doctor on (b)(6) 2020.The issue was not resolved.The patient stopped the ins and would use an external neurostimulator until the scheduled appointment.The patient was alive with injury.Additional information was received reporting that the lead was tested in the operating room, the patient can¿t use the internal stimulation anymore.The patient uses external stimulation.No action or change was planned.
|