• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INTERLINK I.V. ADMINISTRATION SETS WITH CONTROL-A-FLO REGULATOR; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INTERLINK I.V. ADMINISTRATION SETS WITH CONTROL-A-FLO REGULATOR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C6891
Medical Device Problem Code Failure to Prime (1492)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the ¿flow meter¿ of an unspecified quantity, within the ten (10) reported units, of interlink system extension sets with control-a-flo regulators would not prime and had to be disconnected.The nurses stated that ¿the flow meter would flow on open but would not drip if flow rate was turned up¿.This issue was identified during the priming step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Additional information was added to h3, h4, and h6.H10: the actual device was not available; however, five (5) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to the specification.Additionally, pressure and clear passage testing were performed with no issues noted.Pull testing was also performed with no issues noted.Furthermore, the companion samples were functionally tested, and the units worked as expected per the specification.Based on the sample testing result, the units were determined to be conforming product.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK I.V. ADMINISTRATION SETS WITH CONTROL-A-FLO REGULATOR
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10805003
Report Number1416980-2020-06940
Device Sequence Number3680506
Product Code FPA
UDI-Device Identifier00085412002422
UDI-Public(01)00085412002422
Combination Product (Y/N)Y
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K121634
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2C6891
Device Lot NumberDR18I27020
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/12/2020
Supplement Date Received by Manufacturer11/26/2020
Initial Report FDA Received Date11/06/2020
Supplement Report FDA Received Date12/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-