Model Number TDP0005
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, ethnicity, and medical history were not provided.The device lot number is not available / not reported.The expiration date of the device is not known.The initial reporter email was not reported / available.The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.The device manufacture date is not known as the device lot number is not available / not reported.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during an in-office primary balloon sinuplasty procedure, the trudi probe (tdp0005 / lot#: unknown) for the trudi nav system broke; this occurred while using the bending tool.There was no difficulty in removing the probe from the patient¿s body.The broken probe was replaced, and the procedure was resumed to completion.There was no report of any patient adverse event or complication.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that multiple attempts to obtain product for analysis and obtain additional information were unsuccessful.If product is returned or information provided at a later date, the file will be updated accordingly.[conclusion]: the healthcare professional reported that during an in-office primary balloon sinuplasty procedure, the trudi probe (tdp0005 / lot#: unknown) for the trudi nav system broke; this occurred while using the bending tool.There was no difficulty in removing the probe from the patient¿s body.The broken probe was replaced, and the procedure was resumed to completion.There was no report of any patient adverse event or complication.Based on complaint information, the device was not available to be returned for analysis.The device lot number was not available.The manufacturing documentation review could not be performed without the lot number.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the limited information available and without the product available for analysis, the reported customer complaint could not be confirmed.In addition, without a lot number to conduct a manufacturing documentation review, it is not possible to determine if the reported failure could be related to the manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint sample; however, it is possible that clinical and procedural factors may have contributed to the reported issue.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Updated sections: g.3, g.6, h.2, h.3, h.6, and h.10.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TDP0005
Common Device Name
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
ACCLARENT, INC.
31 technology drive
irvine CA 92618
MDR Report Key10805062
Report Number3005172759-2020-00007
Device Sequence Number3931458
Product Code PGW
UDI-Device Identifier10705031464193
UDI-Public10705031464193
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K193453
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTDP0005
Device Catalogue NumberTDP0005
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2020
Supplement Date Received by Manufacturer12/18/2020
Initial Report FDA Received Date11/06/2020
Supplement Report FDA Received Date12/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-