• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HEMODIALYSIS SYSTEM WITHOUT CDX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HEMODIALYSIS SYSTEM WITHOUT CDX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 190858
Device Problems Thermal Decomposition of Device (1071); Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a wire connecting the rocker switch and the power control board that burned and melted.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The biomed did not observe any burning smell, smoke, spark, flame, arcing, or any other visible heat or electrical damage related to the rocker switch and the wire.It was also confirmed that there was no damage observed on any other components, or any other additional issues, associated with the rocker switch and the wire.The biomed replaced the rocker switch and the wire to resolve the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The parts were discarded by the biomed and are not available to be returned to the manufacturer for physical evaluation.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM WITHOUT CDX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
MDR Report Key10806673
MDR Text Key215245241
Report Number2937457-2020-02022
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100903
UDI-Public00840861100903
Combination Product (y/n)N
PMA/PMN Number
K113427
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2020
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number190858
Device Catalogue Number190858
Was Device Available for Evaluation? No
Device AgeMO
Date Manufacturer Received11/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-