SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Model Number 72200568S
Medical Device Problem Codes Protective Measures Problem (3015); Unexpected Shutdown (4019)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/13/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an arthroscopic procedure, the light source suddenly shut off and presented e1, e3 and e13 error.The procedure was successfully completed with the same device with no surgical delay, the error disappeared when rebooting the device.No patient injuries and complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.The complaint of errors and power loss was confirmed.The ballast fan shuts down when the power button is turned on.Then the ballast overheats.Cause of errors and power loss is a defective electronic component.The complaint investigation has concluded the cause of the failure to be a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL LIGHT SOURCE 500XL XENON
Common Device Name
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10806974
Report Number1643264-2020-01718
Device Sequence Number17187837
Product Code FFS
UDI-Device Identifier03596010650504
UDI-Public03596010650504
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K994084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72200568S
Device Catalogue Number72200568S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Initial Date Received by Manufacturer 10/14/2020
Supplement Date Received by Manufacturer12/22/2020
Initial Report FDA Received Date11/08/2020
Supplement Report FDA Received Date12/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-