SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Model Number 72200568S |
Device Problems
Protective Measures Problem (3015); Unexpected Shutdown (4019)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/13/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during an arthroscopic procedure, the light source suddenly shut off and presented e1, e3 and e13 error.The procedure was successfully completed with the same device with no surgical delay, the error disappeared when rebooting the device.No patient injuries and complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.The complaint of errors and power loss was confirmed.The ballast fan shuts down when the power button is turned on.Then the ballast overheats.Cause of errors and power loss is a defective electronic component.The complaint investigation has concluded the cause of the failure to be a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|