|
It was reported that during shoulder surgery, the first qfix mini suture anchor had no implant on the end and the second had a broken drill.Nothing fell inside the patient.A backup device was available to complete the procedure with no delay or other complications.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
H10 h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found no related failures; this failure mode will be trended to assess for any necessary corrective actions.A relationship, if any, between the subject device and the reported event could not be determined.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Internal complaint reference: (b)(4).
|