| Model Number |
977D260 |
| Medical Device Problem Codes |
Device Damaged Prior to Receipt by User (1570); Impedance Problem (2950); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the lead has impedances all the way through on 0-7 so manufacturer representative asked the doctor to reseat the lead in the wireless external neurostimulator (wens) and he did and then every contact was out.So rep then asked the doctor to move the lead in the 8-15 and it was out on every contact as well.Rep then replaced the lead with a new lead being that it was an issue with the lead (out of box) and the new lead worked, issue was resolved. the cause was unknown.Lead was being mailed back.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis (b)(4):analysis information (b)(6) 2020 13:20:17 cst pli# 10 product id# 977d260 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The returned device passed all testing in the laboratory and no anomalies were identified.Concomitant medical products: product id neu_style t_acc lot# unknown, product type accessory.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|