• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MAR NEUT 32IDX62OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MAR NEUT 32IDX62OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1219-32-062
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Noise, Audible (3273)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330); Joint Dislocation (2374); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon reported to nca (bfarm): "hip-tep left side on (b)(6) 2013 in our hospital.Since (b)(6)2020, after a rotational movement while working in the garden, patient was experiencing increasing pain and discomfort in the left hip joint.After this, he experienced squeaking noises in the left joint plus functional restrictments and pain under load.Hip-tep right side in 3/2007; content with this one.The x-ray examination indicated a dislocated inlay of the hip-tep on the left.Revision of left hip-tep on (b)(6) 2020.".
 
Event or Problem Description
On (b)(6) 2020 the patient was revised for left hip tilted, dislocated cup insert (disassociation).The liner and cup was revised.Cup insert appears tilted and rotated, now opening in a lateral direction, resulting in impingement of the head at the lateral acetabular margin.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 health effect clinical code - appropriate term/code not available (e2402) is used to capture joint injury.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d10 concomitant med products.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned device along with the attached x-ray images confirmed a disassociation event.Error in materials or manufacture was not identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : dhr has been reviewed and no deviations or anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 32IDX62OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10808928
Report Number1818910-2020-24169
Device Sequence Number4089853
Product Code LPH
UDI-Device Identifier10603295013952
UDI-Public10603295013952
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K983014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1219-32-062
Device Catalogue Number121932062
Device Lot Number307238
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Initial Date Received by Manufacturer 10/22/2020
Supplement Date Received by Manufacturer11/23/2020
12/01/2020
12/14/2020
01/14/2021
01/19/2021
Initial Report FDA Received Date11/09/2020
Supplement Report FDA Received Date11/30/2020
12/16/2020
12/21/2020
01/18/2021
01/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; CORAIL2 STD SIZE 13; DELTA CER HEAD 12/14 32MM +9; PINNACLE 100 ACET CUP 62MM; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
-
-