• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HSK III SEAL (4.5MM), 5-PACK; CLAMP, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HSK III SEAL (4.5MM), 5-PACK; CLAMP, VASCULAR Back to Search Results
Model Number HST III SEAL (4.5MM), 5-PACK
Device Problem Activation Problem (4042)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/19/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.Device not returned.
 
Event Description
The hospital reported that during preparation for an endoscopic vein harvesting procedure using hst iii seal (4.5mm), 5-pack, they operated according to the procedure, and when deploying, the spring was caught and did not come out.
 
Manufacturer Narrative
Trackwise #: (b)(4).The lot # 25148603 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event Description
The hospital reported that during preparation for an endoscopic vein harvesting procedure using hst iii seal (4.5mm), 5-pack, they operated according to the procedure, and when deploying, the spring was caught and did not come out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HSK III SEAL (4.5MM), 5-PACK
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
MDR Report Key10809255
MDR Text Key215307654
Report Number2242352-2020-00973
Device Sequence Number1
Product Code DXC
UDI-Device Identifier00607567700307
UDI-Public00607567700307
Combination Product (y/n)N
PMA/PMN Number
K130382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/21/2020
Device Model NumberHST III SEAL (4.5MM), 5-PACK
Device Catalogue NumberHS-3045
Device Lot Number25148603
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-