SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP BR (27.29); BR (27.29) IMMUNOASSAY Back to Search Results
Model Number N/A
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer contacted the siemens customer care center (ccc) to report a discordant, falsely elevated advia centaur xp br (ca 27.29) result on a patient sample.A siemens customer service engineer (cse) was sent to the customer site and performed a total service call.The reagent probes (rp1, rp2, and rp3) positions were recalibrated.The reagent probe aspirations, acid and base pump dispensing, sample probe, and functional system checks were all acceptable.A quality control check by the customer and acceptable.The customer has updated their standard operating procedure (sop) to calibrate when replacing the ancillary readypack reagent.The advia centaur xp br instructions for use (10630988_en rev.T, 2020-02) states: "if the quality control results do not fall within the expected values or within the laboratory's established values, do not report results." based on in the information provided, no product non-conformance identified.The customer is operational.The instruction for use (ifu) under the interpretation of results states the following: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." the assay and instrument are performing within specifications.No further evaluation of the device is required.
 
Event or Problem Description
A falsely elevated advia centaur xp br (ca 27.29) result was obtained on a patient sample and it was considered discordant compared to a lower result when retested.The customer observed out of range quality control (qc) results the next morning.The ancillary reagent had been changed the previous afternoon.A br assay recalibration was performed using new reagent and ancillary reagents.Qc results were repeated and within acceptable ranges.The laboratory's medical director reviewed patient results, and a corrected report was issued for this patient.All other previously reported patient results were not questioned by the physician(s).There are no reports that treatment was altered or prescribed or adverse health consequences due to the discordant, falsely elevated advia centaur xp br (ca 27.29) result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP BR (27.29)
Common Device Name
BR (27.29) IMMUNOASSAY
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avenue
tarrytown NY 10591 5097
MDR Report Key10809549
Report Number1219913-2020-00498
Device Sequence Number19836485
Product Code MOI
UDI-Device Identifier00630414196664
UDI-Public00630414196664
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K982680
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/20/2020
Device Model NumberN/A
Device Catalogue Number10334837
Device Lot Number248
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/15/2020
Initial Report FDA Received Date11/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-