| Model Number |
KIT, TMX NFRS STARTERMG/DL |
| Medical Device Problem Codes |
Low Test Results (2458); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Meter and test strips were not returned for evaluation.Most likely underlying root cause: mlc-58: user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Note: manufacturer contacted customer in a follow-up call to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.Product notification letter sent to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results from results obtained of 77mg/dl.The expected non- fasting blood glucose test result range is 127-134mg/dl.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the living room.During the call a back to back blood test was performed by the customer and produced test results of 77mg/dl non-fasting using true metrix meter.The test strip lot manufacturer¿s expiration date is 10/12/2020 and open vial date is undisclosed.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Sections with additional information as of 6-jan-2021: h10: products were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strip lot tested and passed.
|
| |
|
Search Alerts/Recalls
|