• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC TRUE METRIX; SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC TRUE METRIX; SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER Back to Search Results
Model Number KIT, TMX NFRS STARTERMG/DL
Medical Device Problem Codes Low Test Results (2458); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Meter and test strips were not returned for evaluation.Most likely underlying root cause: mlc-58: user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Note: manufacturer contacted customer in a follow-up call to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.Product notification letter sent to contact customer care.
 
Event or Problem Description
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results from results obtained of 77mg/dl.The expected non- fasting blood glucose test result range is 127-134mg/dl.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the living room.During the call a back to back blood test was performed by the customer and produced test results of 77mg/dl non-fasting using true metrix meter.The test strip lot manufacturer¿s expiration date is 10/12/2020 and open vial date is undisclosed.The meter memory was reviewed for previous test result history: (b)(6).
 
Additional Manufacturer Narrative
Sections with additional information as of 6-jan-2021: h10: products were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strip lot tested and passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE METRIX
Common Device Name
SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key10809592
Report Number1000113657-2020-00815
Device Sequence Number3799390
Product Code NBW
UDI-Device Identifier00021292006075
UDI-Public(01)00021292006075(21)
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 01/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberKIT, TMX NFRS STARTERMG/DL
Device Catalogue NumberRE4H01-81
Device Lot NumberMW3567S
Was Device Available for Evaluation? No
Distributor Facility Aware Date10/13/2020
Initial Date Received by Manufacturer 10/13/2020
Supplement Date Received by Manufacturer12/10/2020
Initial Report FDA Received Date11/09/2020
Supplement Report FDA Received Date01/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-