| Catalog Number |
4C8723 |
| Medical Device Problem Codes |
Disconnection (1171); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that two (2) clearlink system y-type blood/solution sets were leaking fluid and blood.It was stated that the sets fell apart where one of the spike connections go into the flow chamber.This issue was identified during preparation prior to the start of the procedure.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to d10, h3, h4 and h6.H10: two (2) actual samples were received for evaluation.Visual inspection was performed using the naked eye which observed that the tubing was separated from the blood filter.Additional inspection was performed which observed that the solvent was not applied uniformly in the circumference of the tube.Dimensional testing was performed and no issues noted.The reported condition was verified.The cause of the condition was due to wrong manual assembly during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|